Abstract
The chromatographic conditions were successfully developed for the separation of Nelfinavir Mesylate using Nevirapine as internal standard by Agilent C18 Column (250mm x 25mm)5µm, flow rate was 1ml/min, mobile phase ratio was Methanol: Phosphate buffer (90:10 v/v), detection wavelength was 251 nm. The Spectroscopic method was done in solvent using mobile phase and the instrument Systronics 1170 with UV win software. The instrument used was Agilent HPLC, Separation module 2695, Uv-Vis detector, Empower-software version 2. The retention time was found to be 5.533 min and 4.350 min for internal standard. The % purity of Nelfinavir Mesylate was found to be 100.02%. The system suitability parameters for Nelfinavir Mesylate such as theoretical plates and tailing factor were found to be 11519, 1.04. The analytical method was validated according to ICH guidelines (ICH, Q2 (R1)). The linearity study of Nelfinavir Mesylate using internal standard was found in concentration range of 0.1µg-100µg and correlation coefficient (r2) was found to be 0.999 respectively, % recovery was found to be 99.96% respectively. %RSD for repeatability and precision was found to be <2.LOD value was 80 ng/mL and LOQ value was found to be 100 ng/mL respectively for Nelfinavir Mesylate.